- Low ODs for the Calibrators
Possible cause Resolution Reagent temperature Use reagents at room temperature. Plate shaker Set the plate shaker to a speed of 600 rpm. Plate sealers Do not use during any of the incubation periods. Wells overdrying Reduce time between wash step and addition of next reagent or use a 96-well washer instead of a strip washer. Substrate Use correct volume of substrate. Plate reading Read plate immediately after adding stop solution. Incubation time Incubate for the recommended times. - Controls Out of Range
Possible cause Resolution Control or calibrator storage Store controls or calibrators in ice during use. Incubation times or temperatures Follow protocol for precise incubation times. Plate washing and drying Do not allow plate to overdry after wash step. Detergent present in the wash Use only deionized water. - Results Drift Across the Plate
Possible cause Resolution Time elapsed while loading the plate Load samples within 20-25 minutes. Use a transfer plate to minimize delays. Have all reagents ready in advance to avoid interruptions. - Poor Precision of Replicates
Possible cause Resolution Plate washing step Ensure that plate is washed 4 times per step and that the plate washer is functioning properly. Pipetting Verify that volumes and pipette calibrations are correct. Plate shaker Verify that plate shaker is set at 600 rpm.
The PLAC® test is currently available at some of the leading clinical reference laboratories in the nation:
Quest Diagnostics Incorporated
www.questdiagnostics.com
1-800-553-5445
LabCorp
www.labcorp.com
1-800-762-4344
Mayo Medical Laboratories
www.mayoreference
services.org/mml
1-800-533-1710
ARUP Laboratories
www.aruplab.com
1-800-522-2787
Berkeley HeartLab, Inc.
www.berkeleyheartlab.com
1-877-4-LIPIDS



